EXTENSION AND EVALUATION OF ICP TORCH LIFE SPAN FOR INDUCTIVELY COUPLED PLASMA EMISSION SPECTROMETRY
Keywords:
Repair of damaged ICP torch, Formation of concentric qurtz tube, Performance evaluation.Abstract
The life span of melted Philips torch has been extended for laboratory use by modifying the damaged Philips torch
and then compared with that ofa standard Philips torch. The construction of the modified torch is simple, easy to assemble
and replace, and is ideally suited for the analysis of sample solutions by inductively coupled plasma spectrometry.
Evaluation of the modified torch reduced cost, usually caused by constant melting of ICP torch and also gave similar
results in terms of precision, sensitivity and detection limits when compared with the conventional new ICPPhilips torch
(type 145).
Published
2022-06-09
How to Cite
BAMIRO, F. O., & JOHN, . . D. L. . (2022). EXTENSION AND EVALUATION OF ICP TORCH LIFE SPAN FOR INDUCTIVELY COUPLED PLASMA EMISSION SPECTROMETRY. Biological Sciences - PJSIR, 37(9), 392–394. Retrieved from https://v2.pjsir.org/index.php/biological-sciences/article/view/2724
Issue
Section
Articles